2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 55
F- 5.0.0 FLOW ERROR ‘1’ OR QDL NOT BETWEEN 720 AND 880 OR BC SWITCH > 400
a) IMPORTANT! CLOSE THE SHUNT DOOR!
b) The following is ESPECIALLY IMPORTANT if the machine was
worked on by someone who may have connected the BALANCING
CHAMBER VALVES (#31 through #38) incorrectly between the
distribution board and their solenoid terminals (Figure right)!
c) TWO (2) possible scenarios 1) or 2) below:
1) IF (and ONLY if) SURE the machine was NOT worked on previously for this Flow problem:
See procedure number F- 5.1.0 (page 56).
2) IF the machine WAS or MAY have been worked on: See parts a THROUGH c below:
a) Leave the machine in Dialysis Program and
DO NOT turn Dialysate Flow “OFF”!
b)
Per Figure 9 below, CAREFULLY trace the wires from EACH valve to its distribution
board position.
c) If (and ONLY if) a wiring error is NOT located see procedure number F- 5.1.0 (page 56). If a
wiring error was located this may be the problem.
Figure 9 – Distribution Board / Balancing Chamber Valves

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals