2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 152
SECTION 2 – “NO WATERALARM
A) ENSURE the water is on! If attached to a PORTABLE RO ENSURE it is on, alarm free, AND NOT in
product divert!
B) ENSURE the incoming water line, between the 2008T and the RO, is connected PROPERLY!
C) Call debug screen 0.
D) Ignoring the TOP Flow Error window, look at
Valve Error (2
nd
window
down). TWO (2) possible scenarios 1) or 2) below:
1) IF (and ONLY if) Valve Error = 0: See procedure number NW- 1.0.0 (page 152).
2) IF Valve Error = 1 longer than two (2) seconds consistently: Refer to the Table below:
If the machine is CURRENTLY in:
Your response
Dialysis Program i.e. connected to
Acid and Bicarb
Proceed to
page 711, Section 26
A Cleaning / Disinfection Program
(Rinse, Heat Disinfect, etc.)
Proceed to
page 207, procedure number CLEAN- 7.0.0
NW- 1.0.0 VALVE ERROR = 0
a) Figure below, ENSURE the Vent Tubing is NOT kinked, INCLUDING the short segment going back
to the Float!
b) Is the machine attached to a PORTABLE RO? TWO (2) possible scenarios 1) or 2) next page:

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals