2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 538
AIR- 1.0.4 AIR SEEN AT L10 / TRACE UPSTREAM FROM CHAMBER #69
Per Table 10 below, ENSURING TMP is between 300 and 360 mmHg, for one (1) minute, watch for air
starting at Location #11 (L11).
Table 10 – Air Checks (2)
Location
Description
If air is seen go to:
If NO AIR is seen the leak is from:
L11
Output Blood Leak Sensor
#8 (see Figure 78, page
538)
L12 (next row)
Valve #24 / Valve #26 mounting O-Ring(s) (to
LOCATE refer to Figure 76, page 536) OR
DiaSafe
®
Filter Tubing connection
L12
Input Blood Leak Sensor #8
(see Figure 78)
L13 (next row) Blood Leak Sensor #8
L13
Output Valve #25 (see
Figure 75, page 535)
L1 (next row) Dialysate Pressure Transducer #9
L1
OUTPUT (to machine)
side of Dialysate Filter
#73
(see Figure 74, page
535)
See (ABOVE) procedure
number AIR- 1.0.2
(
page 536).
Valve #25
Figure 78 – Blood Leak Tubing

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals