2008T Troubleshooting Guide
P/N 490292 Rev. A
AIR- 1.0.4 AIR SEEN AT L10 / TRACE UPSTREAM FROM CHAMBER #69
Per Table 10 below, ENSURING TMP is between 300 and 360 mmHg, for one (1) minute, watch for air
starting at Location #11 (L11).
Table 10 – Air Checks (2)
If NO AIR is seen the leak is from:
L11
Output Blood Leak Sensor
#8 (see Figure 78, page
538)
L12 (next row)
Valve #24 / Valve #26 mounting O-Ring(s) (to
LOCATE refer to Figure 76, page 536) OR
DiaSafe
®
Filter Tubing connection
L12
Input Blood Leak Sensor #8
(see Figure 78)
L13 (next row) Blood Leak Sensor #8
L13
Output Valve #25 (see
Figure 75, page 535)
L1 (next row) Dialysate Pressure Transducer #9
L1
OUTPUT (to machine)
side of Dialysate Filter
#73
(see Figure 74, page
See (ABOVE) procedure
number AIR- 1.0.2
(
page 536).
Valve #25
Figure 78 – Blood Leak Tubing