2008T Troubleshooting Guide
P/N 490292 Rev. A
T- 2.1.6 T5 BETWEEN 2.0 AND 28.0 VOLTS .
a) Call debug screen 0. If debug does not appear press ‘Esc’ then call screen 0.
b) Watching for one (1) minute, if Flow Error EVER = 1 OR a “No Water” alarm EVER appears DO NOT
troubleshoot temperature!
c)
CAUTION! The measurement at the Power Logic board’s twenty-pin X2 cable, at pin 7 is repeated
To avoid damage, TWO (2) precautions:
1) Use the protected red meter lead!
2) Per the Figure below, place the lead horizontally onto pin 7 (BOTTOM row, four
pins from the REAR of the machine). 4.0 volts DC or MORE now?
Yes Pin 7 is more than 4.0 volts. See procedure number T- 2.1.7 (page 290).
No Pin 7 is 3.9 volts OR LESS! See (ABOVE) procedure number T- 2.0.6 (
page 286).