2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 107
F- 9.0.42 MORE THAN 800 ML INPUT FLOW TO VALVE #26 / ISOLATE VALVE #24
a) Return to Calibrate Hydraulics → Flow Pressure.
b) Press ‘Enter’ THEN ENSURE the Flow Motor is rotating!
c) Remove the BLUE dialyzer connector from the shunt door.
d) Measure from the dialyzer connector for one (1) minute!
e) Press ‘Enter’ to return to “Calibrate Hydraulics”.
f) More than 800 ml collected?
Yes More than 800 ml! Good flow from Valve #24! See procedure number F- 9.0.43 (page 108).
No Less than 800 ml! See parts a THROUGH f below.
a) Per the Figure below, remove Valve #24’s INPUT tubing!
b) Direct it into the bucket.
c) Return to Calibrate Hydraulics → Flow Pressure.
d) Press ‘Enter’ THEN ENSURE the Flow Motor is rotating.
e) Measure from the tubing for one (1) minute.
f) Press ‘Enter’ to return to “Calibrate Hydraulics”.
g) More than 800 ml collected?

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals