EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 34
F- 1.0.8 LOADING PRESSURE IS NOT ‘OKAY’ (2)
FIVE (5) possible scenarios 1) or 2) or 3) or 4) or 5) below:
1) IF (and ONLY if) EVER MORE THAN 29 psi: Proceed to
page 36, procedure number F- 1.0.9.
2) IF (and ONLY if) REMAINS LESS THAN 15 psi: A) Remove the gauge THEN again SLAM it into
the RINSE port; B) If still REMAINS less than 15 psi proceed to page 124, procedure number
F- 15.0.0.
3) IF (and ONLY if) PEAKS to between 15 and 22 psi but cycles below 11 psi: See procedure
number F- 1.0.8.11 (page 35).
4) IF (and ONLY if) PEAKS to between 23 and 27 psi but cycles below 11 psi: See procedure
number F- 1.0.8.11 (page 35).
5) IF PEAKS to between 15 and 22 psi AND NEVER cycles below 11 psi: NOTING how many turns,
turn Loading Pressure Valve #65’s nut* CLOCKWISE until if a PEAK pressure of between 23 and 25
psi is achieved. TWO (2) possible scenarios 1) or 2) below:
* To LOCATE Valve #65 refer to Figure 6 (page 22)
1) IF (and ONLY if) between 23 and 25 psi CANNOT be achieved: Return Valve #65’s
screw to its ORIGINAL location then proceed to page 124, procedure number F- 15.0.0.
2) IF between 23 and 25 psi CAN BE achieved: See parts a THROUGH c below:
a) Figure right, if threads ARE visible under Valve #65’s
nut see part b. If no threads are visible is installed either
the wrong spring* is installed OR Valve #65 is bad.
* Refer to Figure 6 (page 22)
b) Call debug screen 0.
c) WITHOUT LOOKING AWAY, watch Flow Error (Top
window) for up to five (5) minutes OR until if it EVER = 1
indicating a Flow Problem. TWO (2) possible scenarios:
1) IF (and ONLY if) Flow Error EVER = 1, even just once: Return to (ABOVE)
procedure number F- 1.0.5 (page 30).
2) IF Flow Error REMAINS = 0: A flow problem is NOT occurring at this time but other
problems may be! Refer to the Table of Contents for whatever other problem may be
occuring.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals