2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 91
Yes BC Switch goes to 897 or more, even if only once! Read before performing! Return to
page 30, procedure number F- 1.0.5 but if you return here, because of no drain flow,
THREE (3) possible bad components (see COMPONENT LIST below):
COMPONENT LIST:
1) Bad Actuator-Test Board
1
OR; 2) Bad actuator cable OR; 3) Multiple intermittent
restricted balancing chamber valves #32, #34, #36 and #38
2
.
No BC Switch is NEVER 897 or more. A Flow Error is not presenting at this time! It may be
extremely intermittent. Return to page 52, procedure number F- 3.8.0.
1
To LOCATE the Actuator-Test board refer to Figure 4A (page 10);
2
To LOCATE these valves refer to Figure 20 (page 92)