2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 11
NOTE B – ACTUATOR-TEST BOARD AND IF bibag EQUIPPED INTERFACE BOARD
A) Unless whatever problem you are troubleshooting will not allow entering Service Mode, if the Actuator-
Test Board is replaced, the Voltage Detector Calibration should be performed! See VOLTAGE
DETECTOR CALIBRATION (page 17).
B) Figure below, if equipped with a bibag Connector the ‘Interface Board’ is piggy backed to the
component side of the Actuator-Test Board. It interfaces the bibag hydraulic components, Actuator-Test
Board and the bibag Distribution Board and Distribution Box 2.
Figure 4C – bibag Interface Board