2008T Troubleshooting Guide
P/N 490292 Rev. A
5) IF from RINSE PORT tubing: Secure tubing connections with tie wraps.
6) IF from Regulator #61 OR Valve #41 (not including the Heat Exchanger): A) Ensure incoming
water pressure is not exceeding 100 psi; B) Check Inlet Water Pressure Regulator per the
CALIBRATION PROCEDURES booklet; C) Check tubing connections tie wraps or replace tubing;
4) If Valve #41 is leaking externally the tubing connection, O-ring, or the valve body may be cracked.
7) IF from Valve #39: A) Check the tubing connection; B) Possible bad O-ring between the valve and
the Hydrochamber; C) The valve body may be cracked.
8) IF from the Acid or Bic pump reinforced (output) tubing: Secure tubing connections with tie wraps.
9) IF between Deaeration Pump #20 and the Hydrochamber including Mixing Chambers #82:
Secure tubing connections with tie wraps.
10) IF between BOTTOM BALANCING CHAMBER VALVES, #35 / #37 and the Hydrochamber:
A) Secure tubing connections with tie wraps; B) Possible bad O-ring between the valve and the
balancing chamber; C) A valve body may be cracked.
11) IF between TOP BALANCING CHAMBER VALVES #32 / #34 and the HEAT EXCHANGER:
A) Secure tubing connections; B) Possible bad O-ring between the valve and the balancing chamber;
C) A valve body may be cracked.
12) Per the Figure below, if the machine is equipped for bibag AND leaking from a bibag component: A)
Secure tubing connections with tie wraps; B) If the component itself is leaking replace it!
Figure 105 – Hydraulics Front View / bibag Components