2008T Troubleshooting Guide
P/N 490292 Rev. A
TMP- 1.0.5 LOADING PRESSURE OKAY / ISOLATE AIR REMOVAL SYSTEM
a) Turn the water OFF!
b) The Deaeration gauge is used next. ENSURE it reads 0
inHg before installing it!
c) Allow a “No Water” alarm to occur THEN, after one (1)
minute, the deaeration motor stops.
d) Figure right, tee the gauge into the Inlet (clear tubing)
side of the Deaeration Pump.
e)
Turn the water on and allow the “No Water”
alarm to go away!
f) Is Deaeration Pressure OKAY? Refer to Appendix A (page 757) for what pressure should be.
Yes Deaeration pressure OKAY! See procedure number TMP- 2.0.0 (page 489).
No Deaeration Pressure in NOT OKAY! ENSURING a “No Water” alarm is NOT presenting,
NOTE this page number, as you will return here, THEN proceed to page 543, SECTION 13
- DEAERATION PROBLEMS.