2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 456
FIL- 1.0.3 DEAERATION PRESSURE OKAY (2)
Because the machine was off, [Conductivity] will be unstable! The following procedure requires three (3)
minutes during when [Conductivity] MUST stabilize to between 13.0 and 14.4 mS!
a) OPEN THE SHUNT DOOR AND LEAVE IT OPEN TILL INSTRUCTED!
b) Call debug screen 0. WITHOUT LOOKING AWAY watch Flow
Error for
three (3) minutes! If EVER = 1, even just once,
indicates a masked Flow Error! TWO (2) possible scenarios:
1) IF (and ONLY if) Flow Error ALWAYS = 0! See procedure number FIL- 1.0.4 (page 456).
2) IF Flow Error EVER = 1: Proceed to
page 23, SECTION 1 - FLOW ERRORS IN DIALYSIS
PROGRAM.
FIL- 1.0.4 ISOLATE ‘OUT OF BYPASS’ CIRCUIT
a) From debug screen 0, Figure right, TWO (2) checks:
CHECK #1: Pre Dialyzer Temperature MUST BE stable*
BETWEEN 35.0 and 39.0° C!
CHECK #2: Pre Dialyzer Conductivity MUST BE stable*
BETWEEN 13.0 and 14.4 mS!
* Stable = NOT changing more than 0.1 per minute! If Temperature is about 37.0° C and
Conductivity about 14.0 mS but either one bounces’ more than 0.3 may indicate a large
hydraulic air leak that must be located and repaired!
b) See procedure number FIL- 1.0.5 (page 456).
FIL- 1.0.5 STABLE CONDUCTIVITY / ISOLATE OLC
A bad On Line Clearance (OLC) component may cause Filling Programs EVEN if OLC is not activated or
used! From debug screen 0:
a)
Figure right, if the POST Dialyzer Cond symbol
appears see part b. If NOT skip to procedure number
FIL- 1.0.6 (page 457).
b) Are PRE AND POST Dialyzer Conductivity within 0.4
mS of each other (Yes or No)?
Yes PRE AND POST Conductivity within 0.4
mS of each other! Proceed to page 459,
procedure number FIL- 2.0.0.
No PRE AND POST Conductivity ARE NOT within 0.4 mS of each other! See procedure number
FIL- 1.0.6 (page 457).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals