EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #353 background imageLoading...
Page #353 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 349
Yes Pump spitting! It primed! See procedure number CO- 1.0.3.6.3 (page 349).
No spitting! There is a problem with the Acid Pump!
CO- 1.0.3.6.3 ACID PUMP SPITTING = IT PRIMED
A) Remove the clamp and reattach the Acid Pump’s output tubing.
B) Connect to the jugs and return to Dialysis Program (“Select
Program” → ‘Dialysis’ → ‘Enter’).
C) Set [Dialysate Flow] to 800 ml/min then allow one (1) minute
D) Call debug screen 0. Figure right, based on Valve #105’s dot color:
1) IF WHITE: See procedure number CO- 1.0.3.6.4 (page 349).
2) IF REMAINING BLUE: Check through the Acid Inlet tubing. If acid is drawn, in one direction only,
towards the machine, return to (ABOVE) procedure number CO- 1.0.3.1 (page 340). If acid is NOT
being drawn there is a problem with the Acid Pump.
CO- 1.0.3.6.4 VALVE #105’S ‘DOT’ IS WHITE
A) Return to Service Mode.
B) Press the Service Mode screen’s ‘Options’ button.
C) Call debug screen 0 and ENSURE Valve #105’s ‘dot’ is now blue!
E) Call debug screen 15 to see Acid Press:
1) IF (and ONLY if) a negative sign (-) appears: Attach a syringe, with its plunger pulled out, to
the RED connector, then push on the plunger attempting to create at least positive 90.
2) IF a negative sign (-) DOES NOT appear: Attach a syringe, with its plunger pushed all the way
in, to the RED connector, then pull on the plunger attempting to create at least negative (-)90
F) Are you able to make Acid Press at least positive OR negative (-)90?
Yes Acid Press goes to at least positive 90 OR negative (-)90! There may still be a problem
with the Acid Pump.
No Could not make Acid Press positive 90 OR negative (-)90! FOUR (4) possible problems:
1) Loose bibag Interface Board ribbon cable (refer to Figure 4C (page 11)) OR; 2) Bad
Pressure Sensor #106 OR; 3) Bad bibag Interface Board OR; 4) Bad connection, from
Pressure Sensor #106, inside the Distribution Box 2 (refer to Figure 4D (page 13)).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals