2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 475
FIL- 5.0.0 AIR SEEN AT CHAMBER #69
Call debug screen 1. Look at FILACT (middle column). TWO (2) possible scenarios:
1) IF (and ONLY if) FILACT = 0: See procedure number FIL- 5.1.0
(page 476).
2) IF FILACT = 1: See parts a THROUGH c below:
a)
Figure below, place one of the resistor plugs, from the FOUR-RESISTOR SET into Air Sensor
Connector’s distribution board position.
b) If (and ONLY if) the plug is placed PROPERLY FILACT, within two(2) minutes = 0 ALWAYS!
c) See procedure number FIL- 5.1.0
(page 476).