2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 403
CO- 8.2.4 AIR = 0 ALWAYS
a) Remove the hydraulics from the cabinet.
b) Per the Figure below, clamp the tubing attached at the top of Valve #101.
c) Reconnect to acid and bicarbonate!
d) From the Home screen, ENSURE [Dialysate Flow] is set at 800 ml/min!
e) Allow fifteen (15) minutes OR until if the external meter’s reading to within +/- 0.4 of TCD?
Yes Within +/- 0.4 of TCD! Remove the clamp from Valve #101 and allow ten (10) minutes. If
measured conductivity becomes high again this confirms a problem with Valve #101. TWO (2)
possible problems: 1) Bad Valve #101 OR 2) Bad bibag Interface Board (refer to Figure 4C
(page 11)).
No NOT within +/- 0.4 of TCD! Remove the clamp then see procedure number CO- 8.3.0 (page
404).