2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 736
V104SC- 2.0.0 BIC PRESS DOES NOT ACHIEVE 400
A) Pull the hydraulics away from the cabinet.
B) Figures below, per the identification (ID) bands on the tubing, ENSURE the bibag Connector tubing
segments have not been reversed or kinked
C)
Leaving the hydraulics out, return to RINSE!
D) Call debug screen 15. If debug does not appear press ‘Esc’ then call screen 15.
E) Watch
Bic Press for one (1) minute. Does it achieve more than POSITIVE 400 now (Yes or No)?
Yes Bic Press achieves POSITIVE 400! A tubing segment may have been kinked! Reinstall the
hydraulics ENSURING the tubing remains unrestricted!
No Bic Press does NOT achieve POSITIVE 400! See procedure number V104SC- 2.0.1 (page
737).