2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 446
EOS- 5.0.0 BOTH PUMP SYMBOLS ARE PINK
a) Turn the machine OFF to prevent damage!
b) Per the Figure below, ENSURE BOTH pumps are installed with their output nozzles (red or blue)
towards the top and the ribbon cables extending from the side. If not, this may be the problem!
c) Swap in a known good Actuator-Test Board* *To LOCATE the board refer to Figure 4A (page 10).
d) If the problem was occurring in a Cleaning Program return to it. If not return to Dialysis Program.
e) Call debug screen 0. If debug does not appear press Esc’ then call screen 0.
f) Allow thirty (30) seconds THEN WITHOUT LOOKING AWAY, watch BOTH pump symbols for two (2)
minutes. Are BOTH still pink?
Yes Both pink! See procedure number EOS- 5.0.1 (page 446).
No Neither pump pink! The previous Actuator-Test Board is bad.
EOS- 5.0.1 ISOLATE THE PUMPS
a) Turn the machine off to prevent damage!
b) Per the Figure above, plug in a known good pump into distribution board position P16, “Conc-P” (Acid
Pump)* ORP16, Bic-P” (bicarbonate pump)*. * Your choice.
c) If the problem was occurring in a Cleaning Program return to it. If not return to Dialysis Program.
d) Call debug screen 0. If debug does not appear press Esc’ then call screen 0.
Part e next page

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals