2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 759
APPENDIX C CALIBRATE IF REPLACED (bibag components)
Components may need to be replaced to repair the machine. The following table
indicates calibrations that must be performed if a bibag component is replaced.
Replaced bibag Component
bibag Technicians Manual
- Conductivity Cell #113
- Not applicable
- Conductivity Cell #117 - Bicarbonate Conductivity Cell
- Temperature Sensor #114 - Not applicable
- Temperature Sensor #116 - Not applicable
- Bicarbonate Pressure Sensor #110 - Pressure Transducers
- Acid Pressure Sensor #106 - Pressure Transducers
- Interface Board (piggy-backed to
the Actuator-Test Board)
- Bicarbonate Conductivity Cell
- Pressure Transducers