Procedure number F- 3.6.0 continued from previous page:
1) IF (and ONLY if) REMAINS = 897 or more: Return to (ABOVE) procedure number
F- 2.0.0 (page 38).
2) IF (and ONLY if) NEVER 897 or more: The Flow Error is not occurring now. Replacing
the Diasafe
®
filter may have fixed the problem HOWEVER, watch for BC Switch for five
(5) more minutes.
3) IF EVER cycles to 897 or more but DOES NOT remain 897 or more: See procedure
number F- 3.8.0 (page 52).
F- 3.8.0 INTERMITTENT FLOW ERRORS
a) It is RECOMMENDED to perform an acid clean as bicarb precipitate may restrict hydraulic pathways
causing intermittent flow errors.
b) Return to Dialysis Program (“Select Program” → Dialysis’ → ‘Enter’).
c)
OPEN THE SHUNT DOOR and LEAVE IT OPEN till instructed!
d) From the Home screen, press the [Dialysate Flow] window.
e) Set [Dialysate Flow] to 800 ml /min and press ‘Enter’.
f) Allow [Conductivity] to increase to at least 13.0 mS.
g) TWO (2) possible TMP window scenarios:
1) IF (and ONLY if) the TMP window is NOT red: See procedure number F- 3.8.1 (page 53).
2) IF the TMP window is red (TMP alarm): See parts a THROUGH c below:
a) Press and release the ‘Reset’ key then immediately press and hold it for three (3) seconds.
b) Allow thirty (30) seconds.
c) If a TMP alarm reoccurs, repeat parts a and b up to twice more BEFORE continuing to
procedure number F- 3.8.1 (page 53).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals