2008T Troubleshooting Guide
P/N 490292 Rev. A
DIASAFE- 5.0.0 ATMP STAYS 300 OR MORE / ISOLATE DIASAFE FILTER #92
a) Figure below, remove FILTER #92 from Valve #28’s tubing.
DiaSafe
®
Filter / Valve #28 / Filter #92
b) Reset ALL alarms.
c) Press the [Test & Options] tab.
d) Press the ‘Diasafe Test’ button then ‘Enter’ then allow the test to finish! TWO (2) possible scenarios:
1) IF (and ONLY if) “Diasafe Test Passed”: FILTER #92 MAY be restricted! Repeat the test
once more as the failure may be intermittent!
2) IF “Diasafe Test Failed”: See procedure number DIASAFE- 6.0.0 (page 633).