2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 135
F- 16.0.9 MORE THAN 130 ML PER MINUTE / ISOLATE FLOAT SWITCH #5
a) This procedure requires up to five (5) FULL minutes to prevent labor-intensive error!
b) Direct the tubing into a bucket for up to five (5) minutes. It will either: 1) Continue at more than
130 ml every minute OR 2) Stop.
1) IF (and ONLY if) flow continues at more than 130 ml every minute! The float is okay! See
procedure number F- 16.2.0 (page 136).
2) IF flow STOPS! TWO (2) possible bad components: 1) Bad Float Switch #5
a
OR; 2) Bad
hydrochamber.
a
To LOCATE Float Switch #5 refer to Figure 28 (page 140)