EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #228 background imageLoading...
Page #228 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 224
TROUBLESHOOTING VALVE ERRORS IN CLEANING PROGRAMS
A) To prevent damage, turn the machine OFF!
B)
Per the Figure below, at the distribution board, THREE (3) checks:
Check #1: ENSURE the ACTUATOR CABLE is plugged in securely AND is not damaged!
Check #2: ENSURE the black PGND is plugged in securely AND shows no sign of burning!
Check #3: ENSURE all Valves are plugged PROPERLY into their distribution board positions
Figure 41 – Distribution Board
B) If a problem was NOT located in part A, see procedure number VEC- 1.0.0 (page 224). If a problem
was located AND REPAIRED return to Heat Disinfect OR Rinse. If VERR (debug screen 1) remains = 0
for five (5) minutes the problem has been eliminated. If VERR does NOT remain = 0 see procedure
number VEC- 1.0.0 (page 224).
VEC- 1.0.0 ISOLATE THE VALVE ERROR / FLOW THROUGH ‘BOB’?
a) Return to Heat Disinfect OR Rinse!
b) Watching for one (1) minute, does the flow indicator’s ’bob’ EVER rise, at least ¼ way up, in the sight
tube?
Yes ‘Bob’ moving! Proceed to
page 230, procedure number VEC- 2.0.0
No ‘Bob’ NOT MOVING! See procedure number VEC- 1.1.0 (page 225).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals