2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 609
LEAKING- 5.0.0 ‘TO DRAIN’ PATH OKAY / ISOLATE VALVE #24
a) A 1000 ml graduated cylinder is required!
b) ENSURE the external flow indicator’s ‘bob’ is at the bottom of the sight tube!
c) Flow from the shunt door and the flow through the indicator will be checked. Parts d through f MUST be
performed QUICKLY so read them BEFORE performing!
d) Place the machine into
RINSE!
e) Remove the RED dialyzer connector from the shunt door but
DO NOT shut the door!
f) Figure right, strong flow, more than 300 ml every thirty (30)
seconds?
Yes More than 300 ml! BEFORE returning the connector
to the shunt and while rinse is still running ENSURE
the external indicator’s ‘bob’ is moving, at least ½ way
up, through the sight tube. See procedure number
LEAKING- 5.1.0 (page 609).
No Not more than 300 ml per 30 seconds! Turn the machine OFF then perform parts a THROUGH c
below:
a) ENSURE no tubing restrictions to and from the DiaSafe
®
filter including inside its housing!
b) Replace the DiaSafe
®
filter with a primed one from another machine*!
*
NOTE! Using a new unprimed filter may cause error!
c) Repeat (ABOVE) procedure number LEAKING- 5.0.0 (page 609). If still not more than 300 ml
per thirty (30) seconds! NOTE
ONLY VALVE #24 will be checked and proceed to page
210, TROUBLESHOOTING A VALVE.
LEAKING 5.1.0 VALVE #24 OKAY / ISOLATE VALVE #25
a) Press the ‘Esc’ key then ‘Enter’ twice to call the Select Program screen.
b) Return the dialyzer connector to the shunt and close the door!
c) Parts d and e MUST be performed quickly! Read them BEFORE performing!
d)
Return to RINSE Program!
e) Is the external flow indicator’s ‘bob’ rising at least ½ way?
Yes ‘Bob’ moving! Valve #25 is okay! See procedure number LEAKING- 5.2.0 (page 610).
No ‘Bob’ NOT moving! TWO (2) possible scenarios 1) or 2) next page:

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals