2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 105
F- 9.0.4 INPUT FLOW PUMP FLOW LESS THAN 800 ML / ISOLATE V26 INPUT FLOW
a) CAUTION! To prevent further damage, Remove the clamps from Valve #43.
b) Obtain an empty bucket.
c)
Figure below, remove Valve #26’s INPUT tubing and direct it into the bucket!
d) Return to Calibrate Hydraulics → Flow Pressure.
e) Press ‘Enter’ then ENSURE the Flow Motor is rotating!
f) Direct Valve #26’s tubing into the graduated cylinder for one (1) minute.
g) Press ‘Enterto return to Calibrate Hydraulics!
h) Reattach Valve #26’s tubing.
i) MORE than 800 ml collected?
Yes MORE than 800 ml! Proceed to
page 107, procedure number F- 9.0.42.
No Less than 800 ml! See procedure number F- 9.0.41 (page 106).