2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 605
LEAKING- 3.2.0 CONTINUE TO ISOLATE LEAK
With the machine OFF! FIVE (5) possible scenarios 1) or 2) or 3) or 4) or 5) below:
1) Figure right, if (and ONLY if) the DiaSafe
®
filter was leaking from its blue
headers: A) ENSURE the packing tabs are not present; B) Replace the filter!
2) Per the Figure below, IF (and ONLY if) Flow Relief Valve #78 was leaking: See
procedure number LEAKING- 3.3.0 (page 606)
3) IF (and ONLY if) the leak was from inside the DiaSafe
®
filter housing:
Inside the housing, tie wrap ALL tubing connections THEN see procedure
number LEAKING- 4.0.0 (page 607).
4) Per the Figure below, IF (and ONLY if) the leak was between the Flow Pump
and BALANCING CHAMBER VALVES #36 and/or #38: See procedure
number LEAKING- 3.3.0 (page 606).
5) All OTHER scenarios: See procedure number LEAKING- 4.0.0 (page 607).