2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 360
CO- 1.0.8.2 CONDUCTIVITY] MORE THAN 14.5 mS / ISOLATE DRAIN STREAM.
a) ENSURE Loading Pressure (Acid Rinse Port gauge) continues to reach a peak more than 22 psi.
b) If a ‘Quick Connector’ (Figure right) is
present at the end of the drain tubing
an ADAPTOR is required.
c) As seen in the Figure (below right), point the drain
tubing opening
UP at 45° holding it no higher
than two (2) feet above the floor. IF POINTED
DOWN GRAVITY FLOW RESULTS IN
ERROR!
d) Watch for ONE (1) FULL MINUTE.
e) Consider BOTH scenarios below AND proceed
accordingly:
Scenario #1:
See Figure A. Approximately 30 ml pulses that stop between each and every
cycle i.e. Pulse → Stop → Pulse → Stop → etc. Reconnect the drain then see
procedure number CO- 1.0.8.3 (page 360).
Figure A
Scenario #2: See Figure B. If a continuous weak stream that LASTS for several seconds or
longer! ENSURING the drain tubing opening was pointed up, proceed to page 66,
procedure number F- 6.2.0.
Figure B
CO- 1.0.8.3 DRAIN FLOW PULSE → STOP / [CONDUCTIVITY] MORE THAN 14.5 mS
A) Return BOTH concentrate connectors to their RINSE ports
B) Close the shunt door.
C) Place the machine into RINSE.
D) Is the external flow indicator’s ‘bob’ moving at least ¼ up in the sight tube?

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals