2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 332
T- 7.8.0 [MONITOR REFERENCE] CHANGES MORE THAN +/- 0.2° C
a) Figure right, place the 34° C (6.808 KΩ)
plug from the FOUR RESISTOR SET into
the 2
nd
distribution board position from
the left “MON-NTC”.
b) TWO (2) possible scenarios ,
based on the screen’s [Monitor
Reference] data box:
1) IF (and ONLY if) [Monitor Reference] continues to change more than 0.2 °C per minute:
THREE (3) possible bad components: 1) Unstable +12 and / or -12 volt DC voltage (i.e. possible
bad Power Logic Board); 2) Bad Sensor Board; 3) Bad Functional Board.
2) IF [Monitor Reference] does DOES NOT change more than 0.2 °C per minute: NTC #3 may be
bad: See parts a AND b below:
a) Replace NTC #3* with a known good! *To LOCATE NTC #3 refer to Figure 42 (page 236).
b) Return to (ABOVE) procedure number T- 7.2.1 (page 323).
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals