2008T Troubleshooting Guide
P/N 490292 Rev. A
CLEAN- 4.0.0 GOOD FLOW PUMP PRESSURE / ISOLATE THE CLEANING PROGRAM
a) Leave the gauge installed ENSURE its tubing is NOT kinked (restricting flow)!
b) Open the shunt door and LEAVE IT OPEN until instructed.
c) Leaving the machine in whatever Program it is CURRENTLY in: 1) Rinse? OR 2) Heat Disinfect?
1) IF (and ONLY if) in RINSE: Proceed to page 202, procedure number CLEAN- 5.0.0.
2) IF in HEAT DISINFECT: See parts a AND b below:
a) Press ‘Esc’ to call the Heat Disinfect
screen (Figure right).
b) Ignoring the Remaining Time window,
does the VERY TOP window,
Remaining Prerinse Time, = 0:00
min:sec?
Yes Remaining Prerinse Time = 0:00!
See procedure number
CLEAN- 4.2.0 (page 193).
No Remaining Prerinse Time DOES
NOT = 0:00! Proceed to page 202,
procedure number CLEAN- 5.0.0.
CLEAN- 4.2.0 REMAINING PRERINSE TIME = 0:00
a) Call debug screen 0. If the debug screens do not appear
press ‘Esc’ then call screen 0.
b) When Remaining Prerinse Time = 0:00 Valve #29’s ‘dot’
(Figure right) REMAINS blue. WITHOUT LOOKING AWAY,
watch it for twenty (20) FULL seconds.
c) TWO (2) possible scenarios:
1) IF (and ONLY if) Valve #29’s ‘dot’ is ALWAYS blue:
See procedure number CLEAN- 4.3.0 (page 194).
2) IF Valve #29’s ‘dot’ is EVER = white: Proceed to
page 202, procedure number CLEAN- 5.0.0.