2008T Troubleshooting Guide
P/N 490292 Rev. A
d) Return to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘Enter’)!
e) Set [Dialysate Flow] to 800 ml/min and press ‘Enter’.
f) See procedure number CO- 8.1.1 (page 400).
CO- 8.1.1 ISOLATE POSSIBLE VERY INTERMITTENT FLOW ERROR
a) Call debug screen 6.
b) WITHOUT LOOKING AWAY, watch BC Switch (middle column) for THREE (3) MINUTES If it EVER =
897 or more, even just once, indicates a masked Flow Error! Is it EVER = 897 or more?
Yes BC Switch = 897 even if just once! Proceed to page 23, SECTION 1 - FLOW ERRORS
IN DIALYSIS PROGRAM.
No BC Switch is NEVER, EVER = 897 or more! See procedure number CO- 8.2.0 (page 401)
LEFT BLANK INTENTIONALLY