2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 116
F- 12.2.0 NEGATIVE 250 MMHG WAS ACHIEVED / PRESSURE TEST
Does meter pressure HOLD, +/- 15 mmHg, for one (1) minute?
Yes Pressure holds! See procedure number F- 12.2.2 (page 116).
No Pressure does DOES NOT hold! Proceed to
page 117, procedure number F- 12.3.0.
F- 12.2.2 PRESSURE HOLDS / VERIFY DIALYSATE PRESSURE
a) ENSURING the external meter is HOLDING negative (-)250 +/- 5 mmHg press ‘Enter’ twice to save the
calibration.
b) Figure right, TWO (2) possible scenarios based
on if an “Operator Error” banner occurs:
1) IF (and ONLY if) Operator Erroroccurred: See procedure number F- 12.4.0 (
page 118).
2) IF “Operator Errordid NOT occur: See parts a THROUGH e below:
a) Turn the machine OFF!
b) Turn the machine on and return to Dialysis Program (“Select Program”‘Dialysis’ →
‘Enter’)!
c) From the Home screen set [Dialysate Flow] to 500 ml/min and press Enter‘.
d) Allow five (5) minutes BEFORE continuing to part e
e) Return to (ABOVE) procedure number
F- 11.0.5 (page 113).
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals