2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 25
F- 1.0.2NO WATERNEVER APPEARS / ISOLATE AIR IN DIALYSATE FUNCTION
A) Figure below, remove the Distribution Board cover.
B) The female Air Sensor’s connector is the
4
th
connector i.e. 5
th
position FROM THE LEFT. If
CBE modified the connector plugs into the CBE boardtwo pins higher than the others!
C) Unplug the female Air Sensors connector but DO NOT remove the ‘CBE board’ under the connector!
D) Leave the Air Sensor distribution board position vacant for now!
E) Call debug screen 0.
F)
Figure right, reading the text box above Chamber
#69, TWO (2) possible scenarios:
1) IF it says Air always: See procedure number
F- 1.0.2.11 (page 26).
2) IF it says “No AirOR toggles between Air and No Air”: See parts a AND b next page: