2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 201
CLEAN- 4.6.33 ISOLATE VALVE #43 POSSIBLY STICKING CLOSED
PDIA is NEVER 6.0 or AND NEVER 0.0 longer than seven (7) seconds!
a) Call debug screen 0 (Figure right).
b) Allow Valve #43’s ‘dot’ to turn blue then white
again.
c) As seen earlier, ENSURE the external Flow
Indicator’s ‘bobis rising at least ¼ way up in
the sight tube!
d) Per the Figure below, unplug the Flow Pump from Distribution Board position “P21-Flow-P”.
e)
ENSURE the FLOW MOTOR shaft stops rotating!
f) Every forty five (45) seconds Valve #43’s dotturns blue for sixteen (16) seconds. WHILE IT IS BLUE,
does the external flow indicator’s ‘bob’ rise at least ¼ way up in the sight tube?
Yes Bob’ moving! Valve #43 is okay! A) Return the Flow Pump’s connector to distribution
board position” P21- Flow-P”; B) ENSURE the Flow Motor rotates; C) If Flow Recirc
Error 1continues to occur you may have made an error somewhere!
No ‘BobNOT moving! ENSURING Loading Pressure (gauge in the rinse port) remains good,
TWO (2) possible bad components: 1) Bad Actuator-Test Board OR 2) Bad Valve #43
(sticking closed).
NOTE! When troubleshooting is complete be sure to return the Flow Pump’s connector to
distribution board; position” P21- Flow-P”;

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals