2008T Troubleshooting Guide
P/N 490292 Rev. A
DIASAFE- 3.0.0 PRESSURE TESTS PASS / DIASAFE
®
TEST
a) Figure right, ENSURE Valve #28 is plugged
PROPERLY into distribution board position
“V28”!
b) From the Home screen, allow [Conductivity]
to stabilize.
c)
Reset ALL alarms!
d) Press the screen’s lower [Test & Options] tab.
e) Press the ‘Diasafe Test’ button.
f) Press ‘Enter’ to start the test then allow it to finish! TWO (2) possible scenarios:
1) IF (and ONLY if) “Diasafe Test Passed”: Repeat the
test once more as the failure may be intermittent! If it
passes again DO NOT continue!
2) IF “Diasafe Test Failed”: a) Call the Home screen.
b) Set [UF GOAL] to 1000 ml; Set [UF Time] to 1:00 hr.
c) Press ‘Enter’.
d) Allow four (4) minutes BEFORE continuing to part e.
e) Using a flashlight, ENSURE no air bubbles flowing through the
external flow indicator’s sight tube.
f) See procedure number DIASAFE- 4.0.0 (page 631).