2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 630
DIASAFE- 3.0.0 PRESSURE TESTS PASS / DIASAFE
®
TEST
a) Figure right, ENSURE Valve #28 is plugged
PROPERLY into distribution board position
“V28”!
b) From the Home screen, allow [Conductivity]
to stabilize.
c)
Reset ALL alarms!
d) Press the screen’s lower [Test & Options] tab.
e) Press the ‘Diasafe Test’ button.
f) Press ‘Enter’ to start the test then allow it to finish! TWO (2) possible scenarios:
1) IF (and ONLY if) “Diasafe Test Passed”: Repeat the
test once more as the failure may be intermittent! If it
passes again DO NOT continue!
2) IF “Diasafe Test Failed”: a) Call the Home screen.
b) Set [UF GOAL] to 1000 ml; Set [UF Time] to 1:00 hr.
c) Press ‘Enter’.
d) Allow four (4) minutes BEFORE continuing to part e.
e) Using a flashlight, ENSURE no air bubbles flowing through the
external flow indicator’s sight tube.
f) See procedure number DIASAFE- 4.0.0 (page 631).
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals