2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 188
CLEAN- 2.1.2 ISOLATE FLOAT SWITCH #5
a) This procedure requires up to four (4) FULL minutes to prevent labor-intensive error.
b) Direct the tubing into the bucket for up to four (4) minutes. Flow will either: 1) Continue at more than
130 ml every minute OR 2) Stop. TWO (2) possible scenarios:
1) IF (and ONLY if) flow continues at 130 ml every minute: Return to (ABOVE) procedure
number CLEAN- 2.0.3 (page 185).
2) IF flow STOPS: TWO (2) possibilities: 1) Bad Float Switch #5* OR; 2) Bad Hydrochamber.
* To LOCATE the Float Switch refer to Figure 28 (page 140)
LEFT BLANK INTENTIONALLY