2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 106
F- 9.0.41 LESS THAN 800 ML TO VALVE #26
a) ENSURE no restrictions or leaks between the DiaSafe
®
filter and the Balancing Chambers
b) ENSURE no restrictions between the DiaSafe
®
filter and Valve #26.
c)
Figure below, at the rear of the DiaSafe
®
filter housing, remove the tubing from Valve #28 (V28) and
direct it into the bucket.
d) Return to Calibrate Hydraulics → Flow Pressure.
e) Press ‘Enter’ and ENSURE the Flow Motor is rotating!
f) Direct the tubing into the graduated cylinder for one (1) minute.
g) Press ‘Enter’ to return to “Calibrate Hydraulics”.
h) More than 800 ml collected?
Yes More than 800 ml collected! The DiaSafe
®
filter is restricted!
No Less than 800 ml collected! This indicates restricted ‘Fresh Side’ Balancing Chamber valve.
See parts through c below:
a) Turn the machine OFF!.
b) Turn the machine on and return to Dialysis Program (Select Program”‘Dialysis’ →
Enter).
c) Proceed to
page 70 procedure number F- 7.0.0

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals