EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #583 background imageLoading...
Page #583 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 579
Figure 96Balancing Chamber Valves
BOTTOM SIDE LEAK PROCEDURE (Valves #35 through #38)
Theory: The following procedures open the top side valves, one at a time, using 24 volts from valve 30’s
distribution board connector. Because the deaeration pump and Flow Pump are running the bottom balancing
chamber valves are pressurized.
NOTE: Needle nose pliers are required (DO NOT use a plastic clamp) to QUICKLY clamp valve #35’s or
valve #37’s white (bottom) tubing IF a leak is seen. Referring to Figure 96 (page 579), locate these valves at
this time (left side, bottom).
a) Plug valve #31 into valve #30’s distribution board position (V30, DRAIN-V”) and watch valve #31’s
nozzle. Is a leak seen?
Yes QUICKLY clamp valve #35’s white tubing. If the leak stops replace valve #35. If not, replace
valve #36.
No Unplug valve #31 and leave it unplugged. See part b.
b) Plug valve #32 into valve #30’s position (V30, DRAIN-V”) and watch valve #32’s nozzle. Is a leak seen?
Yes QUICKLY clamp valve #35’s white tubing. If the leak stops replace valve #35. If not, replace
valve #36.
No Unplug valve #32 and leave it unplugged. See part c.
c) Plug valve #33 into valve #30’s position (V30, DRAIN-V”) and watch valve #33’s nozzle. Is a leak seen?
Yes QUICKLY clamp valve #37’s white tubing. If the leak stops replace valve #37. If not, replace
valve #38.
No Unplug valve #33 and leave it unplugged. See part d.
d) Plug valve #34 into valve #30’s position (V30, DRAIN-V”) and watch valve #34’s nozzle. Is a leak seen?
Yes QUICKLY clamp valve #37 white tubing. If the leak stops replace valve #37. If not, replace
valve #38.
No A balancing chamber valve leak is not obvious using this test. Turn the machine off and
CAREFULLY plug all valves back into the distribution board. Return to Dialysis Program and
perform automated Alarms / Pressure tests.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals