2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 375
CO- 4.1.0 CENTRAL DELIVERY (SDS) CHECK VALVES INSTALLED?
a) Remove the hydraulics from the cabinet!
b) Figure (below), if check valves (with white inserts) are installed in the Acid and / or Bicarb Pump input
and/or output tubing REMOVE them completely and reattach all tubing with straight fittings!
CAUTION! The check valves at the UF PUMP must NEVER be removed.
c) See procedure number CO- 4.2.0 (page 376).
Figure 58 – UF Check Valves