2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 89
F- 7.4.6 ISOLATE FLOW PUMP INPUT
a) Ensure Select Program” remains up.
b) Return the dialyzer connector to the shunt door.
c) Return valve #35 to valve #30’s distribution board position.
d) Figure right, remove the Flow Pump’s clear INPUT tubing
and measure from it for one (1) minute.
e) 800 ml or more collected?
Yes 800 ml OR more! See procedure number
F- 7.4.6.1 (page 89).
No Less than 800 ml! See parts a THROUGH c below:
a) Reconnect the drain AND Flow Pump tubing.
b) Return valves #30 AND #35 to their distribution board positions.
c) TWO (2) checks below:
CHECK #1 Per the Figure below, check for tubing restrictions to and from post-Cell #13.
CHECK #2 There may be a restriction inside Air Removal Chamber #69!
F- 7.4.6.1 800 ML OR MORE
a) Reconnect the drain and Flow Pump tubing.
b) Return valves #30 AND #35 to their distribution board positions.
c) Plug BOTH concentrate connectors into their rinse ports!
d) See procedure number F- 7.5.0 (page 90).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals