2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 527
PHT- 4.0.0 BOTH PRESSURE TESTS PASS
A) Figure right, if the machine is equipped
with a bibag Connector see part B. If NOT
see procedure number PHT- 4.1.0 (page
527).
B) Is a disposable bibag (pouch) attached (Yes
or No)?
Yes Bibag attached! a) Remove the bibag!
b) Plug the bicarb (blue) concentrate connector into
LIQUID
bicarbonate!
c) Close the bibag door FULLY!
d) Call debug screen 0 (Figure
right) to ENSURE the bibag
door is shown closed!
e) See procedure number PHT- 4.1.0 (page 527).
No a) ENSURE the bicarbonate (blue) connecter is attached to
LIQUID bicarbonate.
b) Call debug screen 0 (Figure right) to ENSURE
the bibag door is shown closed!
c) See procedure number PHT- 4.1.0 (page 527).
PHT- 4.1.0 ISOLATE LOADING PRESSURE
a) ENSURE the Loading Pressure gauge (yellow connector) reads 0 psi before inserting it!
b)
SLAM the gauge into the red Acetate/Acid rinse port. *ELSE pressure will not be read correctly
c) Loading Pressure is ‘OKAY’ if it cycles between a PEAK of somewhere between 23 and 25 psi.
d) FOUR (4) possible scenarios 1) or 2) or 3) or 4) below:
1) IF (and ONLY if) pressure REMAINS ‘OKAY’: Proceed to
page 529, procedure number
PHT- 4.2.0.
Scenarios 2 through 4 next page

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals