EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #60 background imageLoading...
Page #60 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 56
F- 5.1.0 BALANCING CHAMBER VALVES WIRED CORRECTLY
a) Call debug screen 6.
b) Watch BC Switch (middle column) for thirty (30) FULL seconds. Does it REMAIN
ALWAYS = 897?
Yes BC Switch ALWAYS = 897! See procedure number F- 5.1.1 (page 56).
No BC Switch IS NOT always = 897! Proceed to
page 57, procedure number F- 5.2.0 (page 57).
F- 5.1.1 BC SWITCH ALWAYS = 897 / ISOLATE PRESSURE SENSOR CONNECTIONS
a) Figure below, CAREFULLY trace BOTH wire harnesses from distribution board connectors “X9,
P-Dial” AND “X10, CFS” to ENSURE they go to the correct (blue) Pressure Transducer.
b) Are BOTH Pressure Transducers connected properly?
Yes BOTH are connected properly! See procedure number F- 5.2.0 (page 57).
No Connect them properly i.e. CFS #10 → “X10, CFS”; Dialysate Pressure #9 → “X9, P-Dial”
and allow one (1) minute. Call debug screen 1 to watch Flow Error for four (4) minutes. If it
remains = 0 this may have fixed the problem. If it EVER = 1, even just once, continue
procedure number F- 5.2.0 (page 57)
Figure 10 Pressure Sensors

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals