2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 108
Yes More than 800 ml! A) Reconnect Valve #24’s tubing; B) NOTE ONLY VALVE #24
will be checked and proceed to page 210, TROUBLESHOOTING A VALVE.
No TWO (2) possibilities: 1) Bad (restricted) DiaSafe
®
filter OR 2) Tubing restriction
between the balancing chambers and Valve #24.
F- 9.0.43 GOOD FLOW FROM VALVE #24
A restriction is indicated in the ‘from (red) Dialyzer line’.
a) ENSURE no tubing restrictions at the DiaSafe
®
filter.
b) Figure BELOW, open Dialysate Filter #73 housing and ENSURE the filter is clean!
c) Figure above touch Valve #25’s black solenoid. Is it warm?
Yes Solenoid warm! TWO possible bad components: 1) Bad Actuator-Test Board
1
; 2) Bad Valve
#25.
1
To LOCATE the board refer to Figure 4A (page 10).
No Solenoid is not warm! NOTE ONLY VALVE #25 will be checked and proceed to page 210,
TROUBLESHOOTING A VALVE

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals