2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 725
SECTION 28 – HEAT EXCHANGER LEAKING EXTERNALLY
A) Per the Figure below, if the HEAT EXCHANGER’S ‘center screws’ and nuts are not tight this may be
problem!
Figure 135 - Hydraulics Front View
B) A psi pressure gauge is required! ENSURE it reads 0 psi before installing it!
C)
IMPORTANT! Turn the water OFF.
D) Per the Figure below, install the gauge at Inlet Pressure Regulator #61.
E) IMPORTANT! Turn the water on!
F) ENSURE NO LEAKS at the gauge!
G) Plug the concentrate connectors into their rinse ports AND place the machine into
RINSE!
Part H next page