2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 58
F- 6.0.0 DRAIN STREAM CHECKS
a) Figure right, if a ‘Quick Connector’ is present at the
end of the ‘to drain’ tubing an ADAPTOR is
required!
b) Figure below, point the drain tube OPENING
UP at
45°, at a level no higher than two (2) feet above the
floor! IF POINTED DOWN GRAVITY FLOW
RESULTS IN ERROR!
c) Watching for one (1) minute, flow MAY occur every nine (9) seconds!
d) Consider ALL FOUR (4) scenarios below AND proceed with the one that BEST MATCHES:
Scenario #1: Referring to Figure A, approximately 30 ml pulses that stops
completely between each and every cycle i.e. Pulse → Stop → Pulse
→ Stop: NOTE this drain stream for later then see procedure number F- 6.1.0
(page 59).
Figure A
Scenario #2: Referring to Figure B, approximately 30 ml Pulse → Stop → Pulse →
‘Dribble’ (a noticeably weaker stream that lasts about two (2) seconds) → etc.
Flow does NOT stop completely between each and every cycle:
NOTE this drain stream (dribble) for later then see procedure number F- 6.1.0
(page 59).
Figure B
Scenarios 3 AND 4 next page

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals