2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 28
F- 1.0.4 FILACT = 1 OR CHAMBER #69 SAYS “AIR”
a) FIGURE BELOW, CAREFULLY place one of the plugs, from the FOUR-RESISTOR SET, into
the Air Sensor’s distribution board position i.e. 5
th
position* from the LEFT.
*
NOTE! If CBE modified the resistor MUST plug into the CBE boardpin for pin!
b) From debug screen 0, based on Chamber #69’s text box now. TWO (2) possible scenarios:
1) IF (and ONLY if) it says “No Air always: See procedure number F- 1.0.5 (page 30).
2) IF is says “Air” OR toggles between Airand No Air”: To avoid unnecessary work, perform
parts a THROUGH e below:
a) BE CERTAIN the resistor plug is placed PROPERLY in the
5
th
position from the left!
b) If CBE modified, using a flashlight, ENSURE the resistor plugs into the CBE board, pin for pin,
especially the TOP pin!
c) If (and ONLY if) the text box now says “No Air” always see procedure number F- 1.0.5 (page
30). If it EVER says “Air” see part d next page!
Parts d and e next page

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals