EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #180 background imageLoading...
Page #180 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 176
CLEAN- 1.2.6 MOTOR CONTINUES TO ROTATE
Recheck Loading Pressure (gauge in Rinse port). THREE (3) possible scenarios 1) or 2) or 3) below:
1) IF (and ONLY if) ALWAYS less than 15 psi: Leaving the machine in the Cleaning Program, see
procedure number CLEAN- 1.5.0 (page 176).
2) IF (and ONLY if) MORE than 25 psi: Figure below, turn Valve #65’s nut counterclockwise (outward) to
adjust to between 23 and 25 psi then proceed
page 190, procedure number CLEAN- 3.0.0.
3) IF cycling to between 15 and 21 psi: Figure below, turn Valve #65’s nut clockwise (inward)
attempting to adjust pressure to a PEAK between 23 and 25 psi. Can you?
Yes Pressure CAN be adjusted to its PEAK! Allow two (2) minutes then if a Flow Error
reoccurs proceed
page 190, procedure number CLEAN- 3.0.0.
No Loading Pressure remains low! Leaving the machine in the Cleaning Program, see
procedure number CLEAN- 1.5.0 (page 176).
CLEAN- 1.5.0 LOADING PRESSURE REMAINS LOW
a) Turn the Heater Breaker Switch OFF!
b) Figure right, unplug the Deaeration Pump from
distribution board position, “P20, DEGAS-P”.
c) Obtain a
1000 ml graduated cylinder AND
the white drip bucket.
d) See procedure number CLEAN- 1.5.1 (page 177).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals