2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 314
Yes LESS than 300 Ω! See procedure number T- 5.0.2 (page 314).
No MORE than 300 Ω! ENSURING the machine was OFF, prior to measuring, see parts a
THROUGH d below:
a) ENSURE the
274 Ω plug, from the TWO-RESISTOR SET, is in the 2
nd
position,
from the left, “MON-NTC”.
b) ENSURE the plug is aligned with the other connectors.
c) Use the
274 Ω plug from a different TWO-RESISTOR SET!
d) Per the Figure below, measure again at pin 12 ENSURING good contact with the pin! Is
the meter’s numeric AND units display less than 300
Ω now (Yes or No)?
Yes LESS than 300 Ω! See procedure number T- 5.0.2 (page 314).
No MORE than 300 Ω! TWO (2) possible bad components: 1) Sensor cable OR
2) Distribution board
T- 5.0.2 LESS THAN 300 Ω
A) Remove the 274 Ω plug.
B) Return the 6.04 KΩ plug, from
the TWO-RESISTOR SET, to the
2
nd
distribution board position from
the left, “MON-NTC”.
C) Return to Dialysis Program (“Select Program” → ‘Dialysis’ → ‘Enter’).
D) Proceed to
page 315, procedure number T- 5.0.3.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals