EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #399 background imageLoading...
Page #399 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 395
CO- 7.0.0 TROUBLESHOOTING COND CELL #7 CIRCUIT
a) IMPORTANT! Turn the machine OFF!
b) Figure right, place the
274 Ω
resistor plug, from the TWO-
RESISTOR SET, into
Conductivity Cell’s #7
distribution board position, “X7,
COND”.
c) Install a conductivity meter into the dialysate lines to measure ‘actual’ conductivity subsequently!
d) Place the machine into Service Mode → Calibrate Sensors → Cond Cells. Follow the screen’s
instructions through: 4. Waiting until the Conductivity value is stable.”
e) ALLOW six (6) minutes before continuing!
f) Verify measured Conductivity is between 13.0 and 14.3!
g) If Cond Cell #7’s reading stabilizes, 5. Conductivity value is stable….” appears?
Yes “5. Conductivity value is stable…” DO NOT continue with the calibration! Cond Cell #7
may be bad. To LOCATE Cell #7 refer to Figure 57 (page 374).
No See procedure number CO- 7.3.0 (page 396).
LEFT BLANK INTENTIONALLY

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals