EasyManuals Logo
Home>Fresenius Medical Care>Medical Equipment>2008T

Fresenius Medical Care 2008T Troubleshoot Manual

Fresenius Medical Care 2008T
768 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #283 background imageLoading...
Page #283 background image
2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 279
T- 1.8.2 VERIFY TEMERATURE CONTROL
This is NOT a routine Temperature calibration. Follow the procedure exactly to avoid error.
a) Install a calibrated Temperature meter (° C) in the dialyzer
lines.
Figure right, flow must MUST be bottom to top.
b) Confirm SHUNT DOOR is closed.
c) From the Service menu select Calibrate Sensors → Temp Control.
d) ‘Sharply’ press ‘Enter’ TWICE to turn the screen’s [TEMP DAC]
data box pale yellow / white.
e) Press [TEMP DAC] to turn it bright yellow.
f) Set [TEMP DAC] to “180”.
g) ‘Sharply’ press ‘Enter’. [TEMP DAC] MUST remain pale yellow/ white i.e. NOT gray.
h) ENSURING flow through the external indicator AND the Heater Switch is on, allow up to five (5)
minutes OR until the external meter reaches 35.5° C or more?
Yes Meter 35.5° C or more! See procedure number T- 1.8.3 (page 279).
No Meter less than 35.5° C! With [TEMP DAC] pale yellow / white, see procedure number
T-
1.8.4
(page 279).
T- 1.8.3 METER MORE THAN 35.5°C / VERIFY [TEMP DAC]
a) Allowing five (5) minutes after each [TEMP DAC] adjustment, continue until the external meter is
between 36.9 and 37.1° C.
b) Does [TEMP DAC] have to be set to 190 or more to bring meter temperature to 36.9 and 37.1° C?
Yes [TEMP DAC] is 190 or more! See procedure number T- 1.8.4 (page 279).
No [TEMP DAC] less than 190! Proceed to
page 327, procedure number T- 7.5.0.
T- 1.8.4 METER LESS THAN 35.5° C / ISOLATE INCOMING WATER TEMPERATURE
If incoming water is less than 20.0° C (from the RO) the machine may not be able to maintain temperature.
a) Using a calibrated temperature meter measure INCOMING WATER from the RO!
b) More than 20.0° C?
Yes More than 20° C! See procedure number T- 1.8.5 (page 280).
No Less than 20.0° C! Incoming water temperature is too low! Once this problem is fixed, most
likely, temperature can be calibrated.

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals