2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 714
DVF- 2.0.0 VERR = 0 / ISOLATE VALVE #24 OPEN CIRCUIT
Call debug screen 2 to see DIAVLO (Figure below). THREE (3) possible scenarios 1) or 2) or 3) below:
1) IF (and ONLY if) DIALVLO REMAINS ALWAYS = 1: Electrical problem with Valve #24. Proceed to
page 716, procedure number DVF- 4.0.0.
2) IF DIAVLO REMAINS ALWAYS = 0: With the shunt door REMAINING open, see procedure number
DVF- 2.1.0 (page 715).
3) IF (and ONLY if) DIAVLO blinks rapidly between 0 and 1: Electrical problem with Valve #24.
Proceed to
page 716, procedure number DVF- 4.0.0.