2008T Troubleshooting Guide
P/N 490292 Rev. A
Bic Press still MORE THAN positive 249 continued:
Yes Bic Press still MORE THAN positive 249. See procedure number P- H.0.5 (page 673).
No Bic Press is now LESS THAN 249. There may have been a bibag Connector tubing kink!
Perform parts a THROUGH c below:
a) Close the bibag Connector door!
b) Place the machine in Dialysis Program or Rinse. If a voltage alarm occurs return to
(ABOVE) procedure number P- H.0.144 (page 671).
c) Reinstall the hydraulics ENSURING the bibag Connector tubing segments remain
unkinked!
HYDRAULICS FRONT VIEW
P- H.0.5 BIC PRESSURE STILL MORE THAN 249
A) Per the Figure above, pull Tube segment #2 off the bibag Connector i.e. from (blue) Pressure Sensor
#110.
B) Continuous flow (more than 5 ml per minute) through Pressure Sensor #110 (Yes or No)?
Yes Continuous flow! THREE (3) possible problems: 1) Loose bibag Interface board
1
ribbon cable OR;
2) Bad Valve #100
2
OR; 3) Bad bibag Interface Board
1
.
1
To LOCATE the Interface board refer to Figure 4C (page 11)
2
To LOCATE Valve #100 refer to the Figure above.
No flow! See procedure number P- H.0.6 (page 674).