2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 673
Bic Press still MORE THAN positive 249 continued:
Yes Bic Press still MORE THAN positive 249. See procedure number P- H.0.5 (page 673).
No Bic Press is now LESS THAN 249. There may have been a bibag Connector tubing kink!
Perform parts a THROUGH c below:
a) Close the bibag Connector door!
b) Place the machine in Dialysis Program or Rinse. If a voltage alarm occurs return to
(ABOVE) procedure number P- H.0.144 (page 671).
c) Reinstall the hydraulics ENSURING the bibag Connector tubing segments remain
unkinked!
HYDRAULICS FRONT VIEW
P- H.0.5 BIC PRESSURE STILL MORE THAN 249
A) Per the Figure above, pull Tube segment #2 off the bibag Connector i.e. from (blue) Pressure Sensor
#110.
B) Continuous flow (more than 5 ml per minute) through Pressure Sensor #110 (Yes or No)?
Yes Continuous flow! THREE (3) possible problems: 1) Loose bibag Interface board
1
ribbon cable OR;
2) Bad Valve #100
2
OR; 3) Bad bibag Interface Board
1
.
1
To LOCATE the Interface board refer to Figure 4C (page 11)
2
To LOCATE Valve #100 refer to the Figure above.
No flow! See procedure number P- H.0.6 (page 674).

Table of Contents

Other manuals for Fresenius Medical Care 2008T

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Fresenius Medical Care 2008T and is the answer not in the manual?

Fresenius Medical Care 2008T Specifications

General IconGeneral
BrandFresenius Medical Care
Model2008T
CategoryMedical Equipment
LanguageEnglish

Related product manuals