2008T Troubleshooting Guide
P/N 490292 Rev. A
Page 489
TMP- 2.0.0 DEAERATION PRESSURE OKAY
a) Turn the water OFF!
b) Figure right, ENSURE the external flow
indicator’s inner tapered tube is NOT
scored (i.e. tube really ‘cloudy’). If
scored, replace it!
c) Allow a “No Water” alarm to occur
THEN, after one (1) minute, the
deaeration motor stops.
d) Remove the Deaeration gauge and
reattach the tubing!
e)
Turn the water on and allow the “No Water” alarm to go away!
f) Remove the hydraulic compartment from the cabinet!
g) Return BOTH concentrate connectors to their rinse ports.
h) Place the machine into RINSE to pressurize the secondary hydraulic circuit!
ENSURE a “No Water” alarm does NOT occur!
i) Unless a leak is immediately seen allow one (1) minute BEFORE continuing to procedure number
TMP- 2.0.1 (page 490).
LEFT BLANK INTENTIONALLY